United Orthopedic Corporation

20 I nstruments Femoral Offset Adaptor 9303-5311-RA Femoral Cutting Guide Cat. No. varies by size 3.7 Offset Femoral Preparation 3.7.1 Offset Sizing and Placement 1. Access the proper ML size with the Femoral Cutting Guide over the Straight Stem Reamer and against the surface of distal femur. 2. Attach the Femoral Offset Adaptor to the Femoral Cutting Guide thus creating 6 degrees valgus angle for left knee or right knee (by referring to the indicator). Then, secure the adaptor with the upper knob. 3. Evaluate the appropriate offset level with Femoral Offset Bushing . Note: The Femoral Cutting Guide can be stabilized on the distal femur by assembling the Cutting Guide Spacer through the backside groove of the Femoral Cutting Guide . Straight Stem Reamer Cat. No. varies by size Femoral Offset Bushing Neutral: 9303-5310 2 mm: 9303-5312 4 mm: 9303-5314 Cutting Guide Spacer 2 mm: 9303-5402 4 mm: 9303-5404 6 mm: 9303-5406 8 mm: 9303-5408

RkJQdWJsaXNoZXIy MTcyNTM=