I nstruments 7 Tibial Sizing Template Cat. No. varies by size Tibial Offset Bushing 2 mm: 9403-5316 4 mm: 9403-5317 Tibial Sizing Template Handle 9403-1203 Offset Bushing Wrench 9403-5333 5 o’clock direction Alignment Rod 9403-2202 2.4 Offset Tibial Trial Preparation 2.4.1 Offset Sizing and Placement 1. If the position of the Tibial Sizing Template is not satisfactory, conduct the offset procedure. 2. Insert the 2 mm or 4 mm Tibial Offset Bushing onto the reamer and use the Offset Bushing Wrench to rotate it until the proper tibial coverage is achieved. Use the Alignment Rod to confirm alignment. 3. Make a note of the number on the offset bushing that lines to the laser mark on the Tibial Sizing Template . (eg. 5 o’clock position shown above)